@article{oai:mie-u.repo.nii.ac.jp:00005736, author = {田口, 寛 and Taguchi, Hiroshi and 若林, 豊 and Wakabayashi, Yutaka and 石原, 則幸 and Ishihara, Noriyuki and 奥村, 克純 and Okumura, Katsuzumi and 嶋林, 幸英 and Shimabayashi, Yoshihide}, journal = {三重大学生物資源学部紀要 = The bulletin of the Faculty of Bioresources, Mie University}, month = {Jan}, note = {application/pdf, A glutamine sensor was constructed by immobilizing L-glutaminase with glutaraldehyde on ISFET (ion selective field effect transistor) to investigate the possibility of determining L-glutamine. The prin- ciple of this sensor is as follows. The output voltage from ISFET, which depends on a very minute pH elevation due to the ammonium ion formation according to the enzymic reaction, is amplified with a differential amplifier of our own assembled. The amplified voltage was processed automatically with an on-line personal computer via GP-IB interface on a real-time basis. The calibration line for L-glutamine was linear from 1mM to 40mM and the relation coefficient was 0.967. The sensitivity of this system to L-glutamine was 0.5mM. The time needed for deter- mining one sample was within 13 min. The substrate specificity of this sensor was strict and L- glutamine was the sole compound responding to this sensor among analogues tested. D-glutamine was completely inactive to this sensor. It is suggested that L-glutanrine can be determined with this newly constructed ISFET-biosensor system with L-glutaminase.}, pages = {71--75}, title = {ISFETバイオセンサシステムによるL-グルタミン定量の可能性の検討}, volume = {7}, year = {1992} }